9.4.11 Air pressure (
BOD Special
operating mode)
Here you can set the air pressure at the measuring location. The setting
range is 500 hPa to 1100 hPa. The default setting is 954 hPa.
The air pressure is directly connected to the height above sea level (the
air pressure influences the equation for the BOD determination). There-
fore, if you change one of these two parameters, the other one is also
changed automatically.
9.4.12 Limit pressure (
Pressure p
operating mode)
Select the limit pressure here, i. e. the pressure difference compared to
the starting pressure (negative or positive pressure) at which the control-
ler takes the relevant sample into a list which is automatically displayed
after the data have been called up.
With this function, the course of the test can be monitored.
Preset:
- Height
:
500 m
To change it:
Confirm with
<RUN/ENTER>
.
Set the required height with
<
><
>
.
Confirm with
<RUN/ENTER>
.
- Back
(
to the
SETTINGS
menu)
RUN/ENTER
SETTINGS
menu:
Use
<
><
>
to select the
-
Air pressure
menu item.
Preset:
- Air pressure
:
965 hPa
To change it:
Confirm with
<RUN/ENTER>
.
Set the required air pressure
with
<
><
>
.
Confirm with
<RUN/ENTER>
.
- Back
(
return to the
SETTINGS
menu)
RUN/ENTER
Summary of Contents for OxiTop Control OC 110
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Page 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Page 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Page 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Page 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Page 101: ......