7.6.3 Restoring data
In case you lost your controller or your controller is defective, the "Restore data“
function facilitates to restore the data of running measurements using a new or
other controller. This means the data are not lost!
To perform this function the memory of the controller must be abso-
lutely empty! The data of each single measuring head are restored
one after the other in a sequence. It is not possible to leave this
function and to continue it afterwards because then the memory
would no longer be empty.
Select the
- Restore data
menu
item.
Select the
- Restore
data
menu item.
Confirm the restoration for one
measuring head and hold the
controller to the measuring
head. Contact selection:
The controller displays the
serial
number of the selected
measuring head.
After confirming again the data
of the measuring head are read
out and stored in the controller.
The Id number of the sample is not restored, it is always
I
999.
RUN/ENTER
RUN/ENTER
RUN/ENTER