background image

Oxi 1970i

Battery / mains operation

19

ba75676e03

05/2012

Battery / mains operation

The plug-in power supply supplies the measuring instrument with low voltage 
(12 V DC). At the same time, the rechargeable battery is charged (approx. 
5 hours). The LoBat display indicator appears when the rechargeable battery 
is nearly empty.

Warning
The line voltage at the operating site must lie within the input voltage range 
of the original plug-in power supply (see T

ECHNICAL

 

DATA

).

Warning
Only use original plug-in power supplies (see T

ECHNICAL

 

DATA

).

2

1

3

Connect the plug (1) to the socket (2) of the measuring instrument.

Connect the original WTW plug-in power supply (3) to an easily 
accessible mains socket.

Summary of Contents for Oxi 1970i

Page 1: ...ucciones de operaci n Portables Sauerstoff Messger t Seite 3 Portable Oxygen Measuring Instrument Page 17 Appareil portable de mesure de l oxyg ne Page 31 Medidor port til para Ox geno P gina 45 RUN E...

Page 2: ...document via internet under http www WTW com Remarque ce mode d emploi Ce mode d emploi contient sous forme compacte la description de toutes les fonctions de base et toutes les informations assurant...

Page 3: ...r l schen 10 Daten bertragen 11 Schreiber Analogausgang 11 Konfigurieren 12 R cksetzen Reset auf Grundeinstellungen 13 Entsorgung 13 Technische Daten 14 Sicherheit Sicherheits hinweise In den einzelne...

Page 4: ...Sensoren erf llen diese Bedingungen Sal pH mV m cm S cm 1 pH nLF TEC Tref25 Tref20 REL2 REL1 Lin K Sal mg l S O 8 8 8 8 8 88 1 C Time Year LoBat Oxi ARng AR RCL Ident Store TP No Day Month Baud Auto...

Page 5: ...cheint wenn der Akku weitgehend entladen ist Achtung Die Netzspannung am Einsatzort muss innerhalb des Eingangs Spannungs bereichs des Original Steckernetzger ts liegen siehe TECHNISCHE DATEN Achtung...

Page 6: ...r Verf gung AutoRange bewirkt dass das Messger t bei berschreiten eines Messbe reichs automatisch in den n chsten Messbereich wechselt Zum Ein Aus schalten der automatischen Messbereichswahl siehe KON...

Page 7: ...ssung z B Winkler Titration nach DIN EN 25813 bzw ISO 5813 Dabei wird die relative Steilheit ber einen Korrekturfaktor an die Vergleichsmessung angepasst Sensorbewertung Nach dem Kalibrieren bewertet...

Page 8: ...l cm 1 K RCL REL2 pH mV m cm S mg l Sal 0 1 7 6 88 C TP Cal O S AR Sauerstoffsensor an das Messger t anschlie en Taste CAL so oft dr cken bis O2 FAC erscheint Sauerstoffsensor in die Kalibrierl sung t...

Page 9: ...Oxi 1970i Kalibrieren 9 ba75676d03 05 2012 Messwerts in der Anzeige...

Page 10: ...tervall mit einstellen An schlie end RUN ENTER dr cken Erneut RUN ENTER dr cken und Identnummer mit eingeben Mit RUN ENTER in den zuletzt ak tiven Messmodus wechseln Das automatische Speichern ist ein...

Page 11: ...2 bestimmt den zeitlichen Abstand zwischen automatischen Daten bertragungen Nach Ablauf des eingestell ten Intervalls wird der aktuelle Datensatz an die Schnittstelle bertragen Zum Einstelllen des be...

Page 12: ...en mit einstellen Auswahl 1 14 999 d Anschlie end Taste RUN ENTER dr cken Automatische Messbereichswahl AutoRange mit aus ein schalten Auswahl YES ein no aus Anschlie end Taste RUN ENTER dr cken Datum...

Page 13: ...Bei Fragen wenden Sie sich bitte an Ihren H ndler Sauerstoff O2 InI Messmodus Sauerstoffkonzentration Automatische Messbereichsum schaltung AutoRange Ein YES relative Steilheit 1 00 Korrekturfaktor d...

Page 14: ...el Metallhydrid NiMH Laufzeit ca 600 Stunden mit einer Akkuladung Steckernetzger t Ladeger t F r alle Steckernetzger te gilt Anschluss max berspannungskategorie II Steckernetzger t mit Euro US UK und...

Page 15: ...0 100 0 mg l 0 2 V 0 1 V 0 01 mg l 0 1 mg l Genauigkeit 0 5 vom Anzeigewert 0 1 S ttigung bzw 0 5 vom Anzeigewert 0 01 mg l Innenwiderstand 5 Ohm Strombegrenzung auf max 0 2 mA Ausgangsstrom Angewend...

Page 16: ...sensor DurOx 325 Gr e Messbereich Aufl sung Konzentration mg l 0 90 0 1 S ttigung 0 600 1 T C 0 50 0 0 1 Genauigkeiten Gr e Genauigkeit 1 Digit Konzentration mg l 0 5 vom Messwert bei Umgebungstempera...

Page 17: ...ing the data storage 24 Transmitting data 25 Recorder Analog output 25 Configuring 26 Resetting to default settings 27 Disposal 27 Technical data 28 Safety Safety instructions The individual chapters...

Page 18: ...fulfill these conditions Sal pH mV m cm S cm 1 pH nLF TEC Tref25 Tref20 REL2 REL1 Lin K Sal mg l S O 8 8 8 8 8 88 1 C Time Year LoBat Oxi ARng AR RCL Ident Store TP No Day Month Baud Auto Cal Status...

Page 19: ...at display indicator appears when the rechargeable battery is nearly empty Warning The line voltage at the operating site must lie within the input voltage range of the original plug in power supply s...

Page 20: ...ge selection There are several measuring ranges available for oxygen measurements If a measuring range is exceeded AutoRange causes the measuring instrument to change automatically to the next measuri...

Page 21: ...ison measurement e g Winkler titration according to DIN EN 25813 or ISO 5813 At the same time the relative slope is adapted to the comparison measurement by a correction factor Sensor evaluation After...

Page 22: ...S mg l Sal 0 1 7 6 88 C TP Cal O S AR Connect the D O sensor to the measuring instrument Press the CAL key repeatedly until O2 FAC appears Immerse the oxygen sensor in the calibrating solution Press...

Page 23: ...Oxi 1970i Calibration procedure 23 ba75676e03 05 2012 If the correction factor deviates from 1 000 the display of the measured value units flashes...

Page 24: ...set the interval with Then press RUN ENTER Press RUN ENTER once more and enter the ID number with Switch to the last active measuring mode with RUN ENTER The automatic save is switched on display Auto...

Page 25: ...l for the data transmission Int 2 determines the chronological interval between automatic data transmissions After the selected interval expires the current data record is transmitted to the interface...

Page 26: ...en measurements with Selection 1 14 999 d Then press the RUN ENTER key Switch the automatic measuring range selection AutoRange on off with Selection YES On no Off Then press the RUN ENTER key Select...

Page 27: ...u have any questions please contact your supplier Oxygen O2 InI Measuring mode D O concentration AutoRange automatic switchover of the measurement range On YES Relative slope 1 00 Correction factor of...

Page 28: ...ry Nickel metal hydride NiMH Operational life approx 600 hours with one charging Plug in power supply unit charging device The following applies to all plug in power supplies Connection max Overvoltag...

Page 29: ...l 0 100 0 mg l 0 2 V 0 1 V 0 01 mg l 0 1 mg l Accuracy 0 5 of display value 0 1 saturation or 0 5 of the display value 0 01 mg l Internal resistance 5 Ohm current limited to max 0 2 mA output current...

Page 30: ...325 Variable Measuring range Resolution Concentration mg l 0 90 0 1 Saturation 0 600 1 T C 0 50 0 0 1 Accuracies Variable Accuracy 1 Digit Concentration mg l 0 5 of measured value at ambient temperatu...

Page 31: ...es 38 Transmission de donn es 39 Enregistreur sortie analogique 39 Configuration 40 Restaurer reset l tat initial 41 Elimination 41 Donn es techniques 42 S curit Consignes de s curit Dans les chapitr...

Page 32: ...issent ces conditions Sal pH mV m cm S cm 1 pH nLF TEC Tref25 Tref20 REL2 REL1 Lin K Sal mg l S O 8 8 8 8 8 88 1 C Time Year LoBat Oxi ARng AR RCL Ident Store TP No Day Month Baud Auto Cal Affichage d...

Page 33: ...fiche lorsque l accu est presque vide Attention La tension du secteur du lieu d utilisation doit se situer dans la plage de tension d entr e du transformateur d alimentation original voir CARACT RISTI...

Page 34: ...ure pour les mesures d oxyg ne AutoRange assure que en cas de d passement d une plage de mesure l appareil de mesure commute automatiquement sur la plage de mesure suivante Pour activer d sactiver la...

Page 35: ...bration en passant par une mesure comparative p ex titration Winkler selon DIN EN 25813 ou ISO 5813 Ici la pente relative est adapt e la mesure comparative par un facteur de correction Evaluation de s...

Page 36: ...S mg l Sal 0 1 7 6 88 C TP Cal O S AR Raccorder la sonde oxyg ne l appareil de mesure Appuyer sur la touche CAL jusqu affichage de O2 FAC Immerger la sonde oxyg ne dans la solution de calibration App...

Page 37: ...Oxi 1970i Calibration 37 ba75676f03 05 2012 Si le facteur de correction d rive de 1 000 l unit de la valeur mesur e de l indication affich e clignote...

Page 38: ...on affich e Int 1 et ajuster l intervalle avec Ensuite appuyer sur RUN ENTER Appuyer nouveau sur RUN ENTER et entrer le num ro d identification avec Commuter avec RUN ENTER sur le mode de mesure actif...

Page 39: ...sion de donn es Int 2 d termine l intervalle de temps entre les transmissions de donn es automatiques Apr s coulement de l intervalle r gl le groupe de donn es actuel est transmis l interface Pour r g...

Page 40: ...d sir pour les mesures d oxyg ne avec Au choix 1 14 999 d Ensuite appuyer sur la touche RUN ENTER Activer d sactiver le choix de la plage de mesure automatique AutoRange avec Au choix YES allum no tei...

Page 41: ...s veuillez vous adresser votre revendeur Oxyg ne O2 InI Mode de mesure Concentration en oxyg ne Commutation de la plage de mesure automatique AutoRange ALLUM YES Pente relative 1 00 Facteur de correct...

Page 42: ...al hydrure NiMH Dur e de service 600 heures environ pour une charge de l accu Transformateur d alimentation chargeur Pour tous les transformateurs d alimentation observer connexion max cat gorie de su...

Page 43: ...V 0 1 V 0 01 mg l 0 1 mg l Pr cision 0 5 de la valeur affich e 0 1 saturation ou 0 5 de la valeur affich e 0 01 mg l R sistance int rieure 5 Ohm Limitation du courant 0 2 mA courant de sortance max R...

Page 44: ...mesure R solution Concentration mg l 0 90 0 1 Saturation 0 600 1 T C 0 50 0 0 1 Degr s de pr cision Grandeur Pr cision 1 digit Concentration mg l 0 5 de la valeur mesur e la temp rature ambiante de 5...

Page 45: ...ia 51 Transferir datos 52 Registrador salida anal gica 52 Configurar 53 Reajustar al valor inicial Reset 54 Eliminaci n desaprovisionamiento 54 Especificaciones t cnicas 55 Seguridad Instrucciones de...

Page 46: ...l pH mV m cm S cm 1 pH nLF TEC Tref25 Tref20 REL2 REL1 Lin K Sal mg l S O 8 8 8 8 8 88 1 C Time Year LoBat Oxi ARng AR RCL Ident Store TP No Day Month Baud Auto Cal Indicaci n del estado actual Indica...

Page 47: ...la bater a est pr cticamente agotada Atenci n El voltaje de la red en el lugar de trabajo debe corresponder al voltaje de entrada del transformador de alimentaci n original vea las ESPECIFICACIONES T...

Page 48: ...geno se disponen de varios rangos de medici n La funci n AutoRange hace que el instrumento cambie autom ticamente al siguiente rango de medici n en el momento de sobrepasar el rango de medici n actual...

Page 49: ...n titulaci n volum trica de Winkler seg n DIN EN 25813 o ISO 5813 En este proceso la pendiente relativa es adaptada a la medici n comparativa por medio de un factor de correcci n Evaluaci n del senso...

Page 50: ...Tref20 ARng 1 Time Year Ident No Day Month Baud Sal cm 1 K RCL REL2 pH mV m cm S mg l Sal 0 1 7 6 88 C TP Cal O S AR Conectar el sensor de ox geno al instrumento Presionar la tecla CAL repetidas veces...

Page 51: ...seado mediante Luego presionar la tecla RUN ENTER Presionar nuevamente RUN ENTER e ingresar el No de ident con Con RUN ENTER cambiar al modo de medici n activado de ltimo La funci n de archivo autom t...

Page 52: ...o entre dos transferencias consecutivas autom ticas de datos Despu s que ha transcurrido el intervalo asignado el conjunto actual de datos es transferido a la interfase Para asignar el intervalo de ti...

Page 53: ...es de ox geno mediante Valores a elecci n 1 14 999 d Luego presionar la tecla RUN ENTER Conectar desconectar la selecci n autom tica del rango de medici n AutoRange con Valores a elecci n YES conect n...

Page 54: ...e a su comerciante Ox geno O2 InI Modo de medici n Concentraci n de ox geno Cambio autom tico del rango de medici n AutoRange Conect YES Pendiente relativa 1 00 Factor de correcci n de la pendiente re...

Page 55: ...NiMH Vida til aprox 600 horas con la bater a completamente cargada Transformador de alimentaci n para conexi n a la red cargador Para todos los transformadores de alimentaci n vale conexi n de sobret...

Page 56: ...actitud 0 5 del valor indicado 0 1 saturaci n porcentual o bien 0 5 del valor indicado 0 01 mg l Resistencia interna 5 Ohm limitaci n de la corriente a max 0 2 mA de la corriente de salida Directivas...

Page 57: ...de medici n Resoluci n Concentraci n mg l 0 90 0 1 Saturaci n 0 600 1 T C 0 50 0 0 1 Exactitudes Dimensi n Exactitud 1 d gito Concentraci n mg l 0 5 del valor medido a temperatura ambiente 5 C 30 C Sa...

Page 58: ...Especificaciones t cnicas Oxi 1970i 58 ba75676s03 05 2012...

Page 59: ...Wissenschaftlich Technische Werkst tten GmbH Dr Karl Slevogt Stra e 1 D 82362 Weilheim Germany Tel 49 0 881 183 0 49 0 881 183 100 Fax 49 0 881 183 420 E Mail Info WTW com Internet http www WTW com...

Reviews: