Multi 9420
Conductivity
ba75964e04
04/2015
83
Calibration
evaluation
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration record
Calibration
/
Cali-
bration data stor-
age
/
Output to
USB flash drive or
printer
-
Outputs the stored calibration data
to the USB-A interface (USB mem-
ory device/USB printer)
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the stored calibration data
to the USB-B interface (PC)
Menu item
Setting/
function
Explanation
Display
Calibration record
Cell constant [cm
-1
]
+++
Within the range
0.450 ... 0.500 cm
-1
Error
Error
Outside the range
0.450 ... 0.500 cm
-1
Error elimination (see section 13
W
CALIBRATION Cond
03.04.2013 07:43:33
TetraCon 925
Ser. no. 09250033
Cell constant 0.476 1/cm
25.0 °C
Sensor
+++