Ion concentration
Multi 9420
54
ba75964e04
04/2015
Calibration
evaluation
After calibrating, the meter automatically evaluates the calibration.
Menu item
Setting/
function
Explanation
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Using
<PRT>
, output the dis-
played calibration record to the
USB-B (PC) interface or the USB-
A (USB printer) interface.
Using
<PRT_>
, output all calibra-
tion records to the USB-B (PC)
interface or the USB-A (USB
printer) interface.
Quit the display with
<ESC>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibra-
tion data storage
/
Output to USB flash
drive or printer
-
Outputs the stored calibration data to
the USB-A interface (USB memory
device/USB printer)
Calibration
/
Calibration data stor-
age
/
Output to RS232/USB
-
Outputs the stored calibration data to
the USB-B interface (PC)
Display
Calibration
record
Magnitude of the slope [mV]
+++
50.0 ... 70.0 or 25.0 ... 35.0
Error
Error
< 50 or > 70
or
< 25 or > 35
Error elimination (see section 13 W