Dissolved oxygen
Multi 9310
54
ba75937e03
12/2011
7.3.6 Calibration data
Displaying the
calibration record
The calibration record of the last calibration is to be found under the menu item,
Calibration
/
Calibration record
. To open it in the measured value display, press
the
<CAL__>
key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration
/
Calibration data storage
/
Display
. To open the
Calibration
menu,
press the
<ENTER>
key in the measured value display.
4.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the AutoRead measurement or accept the calibration
value with
<ENTER>
.
The factor that was set last is displayed.
6.
Using
< > < >
, set the correction factor to adjust the displayed con-
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with
<ENTER>
.
The measuring instrument switches to the measured value display.
The status display
[Factor]
is active.
The calibration record is automatically transmitted to the interface
after calibrating.
O
2
7.92
mg/l
0.973
FDO Comp
22.09.2008
08:00