Multi 9310
Dissolved oxygen
ba75937e03
12/2011
49
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are stable
within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
7.1.2 Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.
Measured parame-
ter
Time interval
Stability in the time interval
D.O. concentration
20 seconds
∆
: better than 0.03 mg/l
D.O. saturation
20 seconds
∆
: better than 0.4 %
D.O. partial pressure
20 seconds
∆
: Better than 0.8 mbar
Temperature
20 seconds
∆
T (° C): Better than 0.02