Multi 9310
pH value
ba75937e03
12/2011
35
Continuing with two-
point calibration
9.
Set the nominal buffer value for the measured temperature with
<
S
S
S
S
><
T
T
T
T
>
.
10. Accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
11. If necessary, finish the calibration procedure as a single-point calibra-
tion with
<M>
.
The calibration record is displayed.
For
single-point calibration
, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.
01.07.2011
08:00
12. Thoroughly rinse the sensor with deionized water.
13. Immerse the pH sensor in buffer solution 2.
14. When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it with
<
S
S
S
S
><
T
T
T
T
>
.
15. Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
16. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>
.
The pH value of the buffer solution is displayed.