Operation
Multi 350i
48
Continuing for three-
point calibration
Repeat the steps 12 to 20 in the same way for the third standard solu-
tion. After finishing the last calibration step, the new calibration values
are displayed as a message (
i
).
You have the following options:
!
Accept the new calibration values with
<MENU/ENTER>
. Subse-
quently, the calibration record is displayed and output to the inter-
face at the same time.
!
To switch to the measured value display without accepting the new
calibration values, press
<M>
or
<ESC>
.
20
Wait for the end of the AutoRead measurement or accept the
calibration value with
<MENU/ENTER>
.
The calibration display for the next standard solution appears.
21
For two-point calibration, select
Exit with 2 point calibration
with
<
▲
> <
▼
>
and confirm with
<MENU/ENTER>
.
The calibration is completed as a two-point calibration.
The new calibration values are displayed as a message (
i
).
You have the following options:
!
Accept the new calibration values with
<MENU/ENTER>
.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
!
To switch to the measured value display without accepting
the new calibration values, press
<M>
or
<ESC>
.
i
Standard = 0.020 mg/l
i
U = 8.4 mV
Terminate AutoRead
ISE <CAL>
i
#2 0.020 mg/l 24 °C
i
Immerse sensor in std. 3
Set temperature:
24 °C
Continue
Exit with 2 point calibration
Set standard:
0.050 mg/l
ISE <CAL>