Multi 350i
Operation
37
Continuing for three-
point calibration
(
Calibration type
TEC
)
14
Thoroughly rinse the electrode with distilled water.
15
Immerse the electrode in buffer solution 3.
16
If necessary, measure the temperature of buffer 3 manually,
then enter and confirm it with
<
▲
> <
▼
>
and
<MENU/ENTER>
in the
Set temperature
setting.
17
Select
Continue
with
<
▲
> <
▼
>
and press
<MENU/ENTER>
.
The buffer is measured.
The measured value is checked for stability (AutoRead).
18
Wait for the end of the AutoRead measurement or
Terminate
AutoRead
and accept the calibration value with
<MENU/ENTER>
.
The new calibration values are displayed as a message (
i
).
You have the following options:
!
Accept the new calibration values with
<MENU/ENTER>
.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
!
To switch to the measured value display without accepting
the new calibration values, press
<M>
or
<ESC>
.
i
Buffer value = 4.010
i
U = 184.0 mV
i
Temperature =
24.8 °C
Terminate AutoRead
pH <CAL>