Operation
Multi 350i
36
Continuing for two-point
calibration
(
Calibration type
TEC
)
8
Thoroughly rinse the electrode with distilled water.
9
Immerse the electrode in buffer solution 2.
10
If the
Set temperature
menu item appears, measure and enter
the temperature of the buffer manually (measurement without
temperature sensor).
11
Select
Continue
with
<
▲
> <
▼
>
and press
<MENU/ENTER>
.
The buffer is measured.
The measured value is checked for stability (AutoRead).
12
Wait for the end of the AutoRead measurement or
Terminate
AutoRead
and accept the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears.
13
For two-point calibration, select
Exit with 2 point calibration
with
<
▲
> <
▼
>
and confirm with
<MENU/ENTER>
.
The calibration is completed as a two-point calibration.
The new calibration values are displayed as a message (
i
).
You have the following options:
!
Accept the new calibration values with
<MENU/ENTER>
.
Subsequently, the calibration record is displayed and output
to the interface at the same time.
!
To switch to the measured value display without accepting
the new calibration values, press
<M>
or
<ESC>
.
i
Buffer value = 10.011
i
U = -177.0 mV
i
Temperature =
2
4.8 °C
Terminate AutoRead
pH <CAL>
i
Buffer recognition TEC
i
Immerse sensor in buffer 3
Exit with 2 point calibration
Continue
pH <CAL>