Multi 350i
Operation
31
Calibration points
Calibration can be performed using one, two or three buffer solutions in
any order (single-point, two-point or three-point calibration). The mea-
suring instrument determines the following values and calculates the
calibration line as follows:
Note
You can display the slope in the units, mV/pH or % (see section 4.3.3).
AutoRead
In calibration, the AutoRead function is automatically activated.
The current AutoRead measurement can be terminated at any time
(accepting the current value).
Calibration record
When finishing a calibration, the new calibration values are displayed
as an informative message (
i
symbol) first. Then you can decide
whether you want to take over these values of the new calibration or
whether you want to continue measuring with the old calibration data.
After accepting the new calibration values the calibration record is dis-
played.
Determined
values
Displayed calibration data
1-point
ASY
!
Asymmetry =
ASY
!
Slope = Nernst slope
(-59.16 mV/pH at 25 °C)
2-point
ASY
SLO
!
Asymmetry =
ASY
!
Slope =
SLO
3-point
ASY
SLO
!
Asymmetry =
ASY
!
Slope =
SLO
The calibration line is calculated
by linear regression.