![wtw MIQ/Blue PS SET Operating Manual Download Page 22](http://html1.mh-extra.com/html/wtw/miq-blue-ps-set/miq-blue-ps-set_operating-manual_1000716022.webp)
Installation
MIQ/Blue PS
3 - 12
ba75573e04
05/2009
3.6
Connecting the voltage supply
The following two sections need only be observed if the
IQ S
ENSOR
N
ET
island should be supplied with power by the MIQ/
Blue PS.
3.6.1
Connection to 100 ... 240 V AC
Warning
If the power supply is connected incorrectly, it may represent a
danger to life from electric shock. Pay attention to the following
points during installation:
z
The MIQ/Blue PS may only be connected by a trained
electrician.
z
The connection of the MIQ/Blue PS to the power supply may
only be carried out when it is not carrying any voltage.
z
The power supply must fulfill the specifications given on the
nameplate and in chapter 6 T
.
z
When installed in a building, a switch or power switch must be
provided as an interrupt facility for the MIQ/Blue PS.
The interrupt facility must
– be installed in the vicinity of the MIQ/Blue PS, easily
accessible by the user, and
– be labeled as an interrupt facility for the MIQ/Blue PS.
z
After it has been installed, the MIQ/Blue PS may only be opened
if the mains voltage has been switched off beforehand.
Materials required
z
Wire end sleeves, suitable for the power line, with suitable crimping
tool
z
1 x cable gland, matched to the cable diameter (see section 3.5 on
page 3-10).
Tools
z
Cable stripping knife
z
Wire stripper
z
Phillips screw driver
z
Small screw driver.
Summary of Contents for MIQ/Blue PS SET
Page 30: ...Installation MIQ Blue PS 3 20 ba75573e04 05 2009...
Page 32: ...Maintenance and cleaning MIQ Blue PS 4 2 ba75573e04 05 2009...
Page 38: ...Technical data MIQ Blue PS 6 4 ba75573e04 05 2009...
Page 44: ...Appendix Installation instructions for several radio links MIQ Blue PS 7 6 ba75573e04 05 2009...
Page 45: ......