Messages and log book
The IQ S
ENSOR
N
ET
continuously monitors the status of the entire sys
-
tem. If the IQ S
ENSOR
N
ET
identifies system changes, a message
appears. New messages can be recognized by the flashing information
symbol or error symbol on the display.
All messages are recorded in the log book.
4.6.1 Message types
The system differentiates two types of messages:
Errors
h
Indicates a critical status in the system or an individual system com
-
ponent that requires immediate action.
In the case of new error messages, the error symbol flashes on the
display.
Information
i
Information that does not require immediate action.
In the case of new information, the info symbol flashes on the dis
-
play.
4.6.2 Log book
The log book is a list with all the messages from all modules. The log
book contains up to 1000 entries. If there are more than 1000 mes
-
sages, the oldest entries are deleted.
New messages can be recognized by the flashing information symbol
or error symbol. The log book always contains the latest message in the
first position. They do not yet have a tick in the status field.
The flashing of the info or error symbol only stops after all detailed mes
-
sage texts in the log book have been opened and marked with a tick
(
✔
) (see section 4.6.3).
In the case of errors, immediately open the detailed mes
-
sage text in the log book and perform the recommended
actions. If the actions have been performed, mark the mes
-
sage as read (see section 4.6.3).
Summary of Contents for IQ SENSOR NET System 2020 XT
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 8: ...Contents System 2020 XT CONNECT 0 6 ba75970e01 10 2012...
Page 24: ...Overview System 2020 XT CONNECT 1 16 ba75970e01 10 2012...
Page 28: ...Safety System 2020 XT CONNECT 2 4 ba75970e01 10 2012...
Page 123: ...System 2020 XT CONNECT Settings setup 5 11 ba75970e01 10 2012 Fig 5 10 Edit list of sensors...
Page 138: ...Settings setup System 2020 XT CONNECT 5 26 ba75970e01 10 2012...
Page 152: ...Network connection System 2020 XT CONNECT 6 14 ba75970e01 10 2012...
Page 178: ...PROFIBUS connection option System 2020 XT CONNECT 8 12 ba75970e01 10 2012...
Page 180: ...Maintenance and cleaning System 2020 XT CONNECT 9 2 ba75970e01 10 2012...
Page 194: ...What to do if System 2020 XT CONNECT 10 14 ba75970e01 10 2012...
Page 204: ...Accessories and options System 2020 XT CONNECT 12 2 ba75970e01 10 2012...
Page 212: ...Lists IQ SENSOR NET System 2020 XT CONNECT 13 8 ba75970e01 10 2012...
Page 215: ......