Turbidity measurement (VisoTurb
®
900-P)
ba75937e10
08/2021
67
Continuing with
three-point calibra-
tion
Repeat the steps 11 to 15 with the third standard solution. The new calibration
values are displayed after the last calibration step was completed.
9.2.5 Calibration data
Displaying the cali-
bration data
Calibration
The calibration record of the last calibration is available in the menu
<ENTER>
/
Calibration record
. To open it in the measured value display, press
the
<CAL__>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in the
measured value display, press the
<ENTER>
key.
16.
If necessary, terminate the calibration as a two-point calibration with
<M>
.
The new calibration values are displayed.
or
Continue with three-point calibration.
TRB
124.0
01.07.2011
08:00
FNU
1010.0 FNU
Cal
Standard
3