![wtw inoLab 740 Operating Manual Download Page 78](http://html1.mh-extra.com/html/wtw/inolab-740/inolab-740_operating-manual_1000561078.webp)
pH
78
6.2.1
Buffer sets
Before each calibration, it is necessary to inspect the
selected calibration procedure and the buffer sets used. You
can use the following buffer sets at the specified
temperatures for an automatic calibration.
Buffer set
Menu
designati
on
pH
values
at
temp
WTW technical buffer
solutions
Buffer
TEC
2.00
4.01
7.00
10.00
25 °C
Standard buffer solutions
according to DIN 19266
Buffer
DIN
1.679
4.006
6.865
9.180
12.454
25 °C
Merck company
Titrisole buffers and buffer
solutions
Riedel de Haen company
Fixanale buffers and buffer
solutions
Buffer
Mrk1
2.00
4.00
7.00
9.00
12.00
20 °C
Merck company
Titrisole buffers and buffer
solutions
Riedel de Haen company
Fixanale buffers and buffer
solutions
Buffer
Mrk2
1.00
3.00
6.00
8.00
10.00
13.00
20 °C
Merck company ready-made
buffer solutions
Buffer
Mrk3
4.66
6.88
9.22
20 °C
Summary of Contents for inoLab 740
Page 18: ...Safety 18...
Page 27: ...Commissioning 27...
Page 32: ...Commissioning 32...
Page 50: ...Basic principles of operation 50...
Page 92: ...pH 92...
Page 134: ...Ion concentration 134...
Page 148: ...Dissolved oxygen D O 148...
Page 174: ...BOD 174...
Page 232: ...Data output 232...
Page 236: ...Maintenance cleaning disposal 236...
Page 270: ...Lists 270...
Page 274: ...Lists 274...