Dissolved oxygen (D. O.)
142
8.2.1
Settings
Calibration interval
The calibration interval is used to determine after how many
days you want to be reminded that a D. O. calibration is due.
8.2.2
Calibration
Proceed as follows to calibrate the measuring system:
Preparatory
activities
Note
The sponge in the air calibration vessel must be moist (not
wet). Follow the instructions in the OxiCal
®
-SL operating
manual.
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Cal
menu.
3
Select the
Cal Int. ...
menu item with
82
.
4
Confirm the
Cal Int. ...
menu item with
g
.
The input field for the calibration interval appears.
5
Option:
Select a digit with
46
.
6
Set the required interval (days) with
82
.
7
Confirm the calibration interval entered with
g
.
8
Use
[ESC]
to leave the input field.
1
Connect the D. O. probe to the measuring module.
2
Keep the OxiCal
®
-SL air calibration vessel ready.