Ion concentration
112
7.4
Measuring the ion concentration
Preparatory
activities
Perform the following preparatory activities when you want
to measure the concentration using ion-sensitive
electrodes:
Note
Incorrect calibration of ion-sensitive electrodes will result in
incorrect measured values. Calibrate regularly before
measuring.
Temperature probe
The temperature measurement is absolutely essential for a
reproducible measurement. You can perform concentration
measurements using an NTC 30 or Pt1000 external
temperature probe using the integrated temperature probe
of another probe, or without a temperature probe. The
temperature probe that is used is indicated on the display by
v
,
y
, or
z
.
Note
The measuring system automatically recognizes the type of
the temperature probe used. If the measurement is made
without a temperature probe, proceed as follows:
1
Connect the electrode to the measuring module.
2
Adjust the temperature of the solutions, or measure
the temperature if the measurement is made without
a temperature probe.
3
Select the
ISE
measuring mode.
4
Calibrate or check the measuring system using the
ion-sensitive electrode.
1
Measure the current temperature of the test sample.
2
Adjust the temperature of the test sample and the
calibration standards to the same temperature
(maximum deviation: ± 2 °C).
Summary of Contents for inoLab 740
Page 18: ...Safety 18...
Page 27: ...Commissioning 27...
Page 32: ...Commissioning 32...
Page 50: ...Basic principles of operation 50...
Page 92: ...pH 92...
Page 134: ...Ion concentration 134...
Page 148: ...Dissolved oxygen D O 148...
Page 174: ...BOD 174...
Page 232: ...Data output 232...
Page 236: ...Maintenance cleaning disposal 236...
Page 270: ...Lists 270...
Page 274: ...Lists 274...