System 182
Relay outputs
6 - 15
ba75503e07
04/2012
6.4.2
System monitoring
Function
The
System monitoring
enables monitoring system errors.
To set up the
System monitoring
function for a relay output you have to
select the
System
option when linking the relay output (see section
It can be used to monitor the following system errors:
Settings
Collective error
message
In the case of a
Collective error
message, the relay opens if one of the
following malfunctions occurs:
One of the sensors properly registered at the controller does not
supply a valid main measured value
One of the sensors properly registered at the controller does not
supply a valid secondary measured value
In any case, the relay remains open for 10 seconds and only closes
when the malfunction is no longer present.
In the following cases, the relay does not open despite an invalid
measured value:
The sensor is being calibrated
The sensor is in the maintenance condition
The sensor is being cleaned with the aid of a valve module in the
system (compressed air operated cleaning system).
Settings
Selection
Explanations
Power failure
On
Off
The
Power failure
On
function
monitors the supply voltage in
the
DIQ/S 182-PR. If the voltage
falls below a critical value, the
relay switches.
Collective error
On
Off
The
Collective error
function simultaneously
monitors the proper function of
all sensors. (for details see
below this table)
Action
Open
For all functions of the
System
monitoring
, the relay action is
set to
Open
.
Summary of Contents for DIQ/S 182-PR
Page 6: ...Contents System 182 0 4 ba75503e07 04 2012 ...
Page 10: ...Overview System 182 1 4 ba75503e07 04 2012 ...
Page 14: ...Safety instructions System 182 2 4 ba75503e07 04 2012 ...
Page 78: ...Operation System 182 4 18 ba75503e07 04 2012 ...
Page 84: ...The SETTINGS menu System 182 5 6 ba75503e07 04 2012 ...
Page 110: ...Relay outputs System 182 6 26 ba75503e07 04 2012 ...
Page 122: ...PROFIBUS connection System 182 7 12 ba75503e07 04 2012 ...
Page 124: ...Maintenance and cleaning System 182 8 2 ba75503e07 04 2012 ...
Page 130: ...What to do if System 182 9 6 ba75503e07 04 2012 ...
Page 142: ...Accessories and options System 182 11 2 ba75503e07 04 2012 ...
Page 146: ...Appendix store separately if required System 182 13 2 ba75503e07 04 2012 ...