Relay outputs
System 182
6 - 20
ba75503e07
04/2012
6.4.7
Sensor-controlled cleaning
With the
Sensor controlled
function, the relay is controlled by a linked
sensor.
Requirements
Controller version from 2.80
Sensor that transmits signals to trigger a cleaning cycle, e.g. UV/VIS
sensor
Settings
Note
Settings of the cleaning process can be done in the
SETTINGS
menu
of the respective sensor.
Setting
Selection/Values
Explanation
Pulse length
Automatic
The duration of the air
cleaning process is
programmed in the
sensor.
The relay automati-
cally takes over the
cleaning duration
from the sensor.
0,5 s
1 s
2 s
3 s
The relay finishes the
cleaning after the
selected interval.
Summary of Contents for DIQ/S 182-PR
Page 6: ...Contents System 182 0 4 ba75503e07 04 2012 ...
Page 10: ...Overview System 182 1 4 ba75503e07 04 2012 ...
Page 14: ...Safety instructions System 182 2 4 ba75503e07 04 2012 ...
Page 78: ...Operation System 182 4 18 ba75503e07 04 2012 ...
Page 84: ...The SETTINGS menu System 182 5 6 ba75503e07 04 2012 ...
Page 110: ...Relay outputs System 182 6 26 ba75503e07 04 2012 ...
Page 122: ...PROFIBUS connection System 182 7 12 ba75503e07 04 2012 ...
Page 124: ...Maintenance and cleaning System 182 8 2 ba75503e07 04 2012 ...
Page 130: ...What to do if System 182 9 6 ba75503e07 04 2012 ...
Page 142: ...Accessories and options System 182 11 2 ba75503e07 04 2012 ...
Page 146: ...Appendix store separately if required System 182 13 2 ba75503e07 04 2012 ...