System 182
Relay outputs
6 - 7
ba75504e07
04/2012
6.3.3
Proportional output
In the case of proportional output, a relay switches cyclically on and off
in a defined measured value range (proportional range). At the same
time, the relay switches with a:
duration of operation that corresponds to the measured value
(pulse-width output, see page 6-9) or
switching frequency (frequency output, see page 6-10).
Proportional outputs can be used in the following way:
Output with one relay:
An output range is defined with a
Start value
and an
End value
. No
output takes place above and below the output range
(see page 6-8).
Output with two relays:
An output range is defined for each relay with a
Start value
and an
End value
. One relay outputs in the upper output range and a further
relay in the lower output range (see page 6-8).