Installation
System 182
3 - 44
ba75504e07
04/2012
3.11.2 Connecting two sensors with compressed air cleaning
Connection scheme of
valve control
Example:
Relay 1 controls the cleaning of sensor 1.
Relay 2 controls the cleaning of sensor 2.
Variant: Relay 1 controls the cleaning of both sensors. Cleaning of both
sensors is carried out with the same settings.
Warning
In this configuration, the free switching contact (here: R3) may be
used to switch SELV voltages only.
Warning
No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages. Always cut off any wires
that are not in use as closely as possible to the cable gland.
DIQ/CHV 1
SACIQ
SACIQ
Valve control line 1
Valve control line 2
Air
Air
DIQ/CHV 2
Sensor 2
Sensor 1
Max. total cable length
SNCIQ(/UG) and SACIQ = 250 m
DIQ/S 182-MOD(/24V)
M
C
S
ESC
OK
Summary of Contents for DIQ/S 182-MOD
Page 6: ...Contents System 182 0 4 ba75504e07 04 2012...
Page 10: ...Overview System 182 1 4 ba75504e07 04 2012...
Page 14: ...Safety instructions System 182 2 4 ba75504e07 04 2012...
Page 78: ...Operation System 182 4 18 ba75504e07 04 2012...
Page 84: ...The SETTINGS menu System 182 5 6 ba75504e07 04 2012...
Page 110: ...Relay outputs System 182 6 26 ba75504e07 04 2012...
Page 126: ...Maintenance and cleaning System 182 8 2 ba75504e07 04 2012...
Page 132: ...What to do if System 182 9 6 ba75504e07 04 2012...
Page 144: ...Accessories and options System 182 11 2 ba75504e07 04 2012...
Page 148: ...Appendix store separately if required System 182 13 2 ba75504e07 04 2012...
Page 149: ......