Putting the sensor into
operation after
calibrating
4.6
Passwords
You can assign and activate two passwords for the DIQ/S 181.
Settings password
protects all settings in the
SETTINGS
menu.
If the password is active, all settings can be viewed but not changed.
The password prompt appears on leaving a setting table with the
Save and quit
command. Resetting the system configuration to
default is also protected.
Calibration password
protects against unauthorized calibration.
If this password is active, the password query appears on calling up
calibration with
<C>
. If no valid password is entered, calibration is
blocked.
Configuring
passwords
Activating/deactivating and changing passwords is carried out in the
SETTINGS
->
System
menu.
Fig. 4-11 SYSTEM SETTINGS menu
Operating notes:
The settings password is configured with the
Settings
a
/
z
menu
5
Confirm with
<OK>
. The display returns to the measured value
display (the measured value flashes as the sensor is still in the
maintenance condition).
6
If calibration was successful, immerse the sensor in the test
sample.
7
Wait for a stable measured value.
8
Switch off the maintenance condition (press
<OK>
three
times). The measured value has stopped flashing.
If calibration was not successful, this is indicated by "----"
on the display and a corresponding message with remedial
actions appears in the log book. Follow the instructions and
repeat calibration.
Summary of Contents for DIQ/S 181
Page 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Page 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Page 103: ......