Automatic switch-on of
the maintenance
condition
The maintenance condition is automatically activated
during calibration. After calibration the sensor remains in the
maintenance condition until the maintenance condition is switched
off manually (see section 4.5)
during a compressed air cleaning cycle. After the cleaning cycle is
finished the maintenance condition is automatically switched off
again.
4.4.3
Log book /
Sensor status
S01
In the menu
SENSOR STATUS S01
you can query the following infor-
mation on the connected sensor:
Type and series number
Software status
Error and info messages ("log book")
Data of the last calibration (only for a sensor that can be calibrated)
Fig. 4-8
SENSOR STATUS S01
Log book
The log book is a list of messages related to the connected sensor. If
there is a new message the error symbol in the header of the measured
value display flashes:
The flashing of the error symbol only stops after all detailed message
texts in the log book have been opened and marked with a checkmark
(
). The operating sequence to do this is described on Seite 55.
Calibration data
You can view the data of the last calibration of the sensor under the
Cal
menu item. With a sensor that was not yet calibrated, this menu item is
missing. The symbol in the
Cal
line quotes the validity state:
h
Flashes if a new or unacknowledged error message is present
in the log book that requires immediate action.
The error symbol also flashes if the operational voltage or
communication are faulty. You can check this in the
PROP-
ERTIES /
SYSTEM STATUS
menu (see section 4.4.1).
Summary of Contents for DIQ/S 181
Page 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Page 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Page 103: ......