Cond 3310 IDS
What to do if...
ba77051e04
07/2017
47
10 What to do if...
10.1 Conductivity
Error message
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
10.2 General information
Sensor symbol
flashes
Display
Meter does not react
to keystroke
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– Measured value outside the
measuring range
– Use suitable IDS conductivity sensor
Cause
Remedy
– Sensor contaminated
– Clean the sensor and replace it if
necessary
– Calibration solution not suitable
– Check the calibration solutions
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring system
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see section
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<ENTER>
and
<On/Off>
key simultaneously