Cond 1970i
Operation
43
ba75692e02
07/2009
4.6.2
Recorder (analog output)
You can transmit data to a recorder via the analog output. Connect the
analog output to the recorder via the AK323 interface cable.
The data output automatically switches to
Recorder output.
Socket assignment
Note
The analog output is activated automatically in the cable by connecting
2 and 3.
The output on the analog output corresponds to the value shown on the
display.
Signal range
The signal range of the analog output depends on the measured vari-
able and the measuring range:
Conductivity
Salinity
1 free
2 Plug coding
3 Ground
4 Analog output
1
2
3
4
RS 232
REC
(internal resistance < 5 Ohm)
Measuring range
Voltage
Resolution
0.000 ... 1.999 µS/cm
0 ... 1999 mV
1 mV
0.00 ... 19.99 µS/cm
0 ... 1999 mV
1 mV
0.0 ... 199.9 µS/cm
0 ... 1999 mV
1 mV
0 ... 1999 µS/cm
0 ... 1999 mV
1 mV
0.00 ... 19.99 mS/cm
0 ... 1999 mV
1 mV
0.0 ... 199.9 mS/cm
0 ... 1999 mV
1 mV
0 ... 500 mS/cm
0 ... 500 mV
1 mV
Measuring range
Voltage
Resolution
0 ... 70.0
0 ... 700 mV
1 mV