AmmoLyt
®
Plus
700 IQ
What to do if ...
8 - 1
ba75651e03
08/2009
8
What to do if ...
8.1
Interpretation of the drift voltage
The drift voltage is influenced by the potential levels of the
measurement electrode and reference electrode. If the potential levels
shift, e.g. caused by aging, both parts can move into the same direction
or into opposite directions. The comparison of the drift voltages of two
or three electrodes enables to draw conclusions concerning the state
of individual electrodes if a matrix adjustment or calibration resulted in
an invalid or considerably deviating drift voltage.
Assessment aids
View the calibration history of the combination electrodes. In the
following cases, assessment is relatively easy:
z
If the drift voltages of all combination electrodes show the same
trend, i. e. the drift voltages have shifted in the same direction
(positive or negative) by approximately the same amount, this
indicates that the potential level of the reference electrode has
shifted. The reference electrode is possibly highly contaminated or
at the end of its service life.
z
If the drift voltages of all combination electrodes do not show a trend
but shift by different amounts into different directions, the potential
level of the reference electrode has not changed considerably. The
invalid drift voltage is probably caused by a defective measuring
electrode. The measuring electrode is possibly highly contaminated
or at the end of its service life.
Summary of Contents for AmmoLyt Plus 700 IQ
Page 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Page 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Page 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Page 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Page 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...