AmmoLyt
®
Plus
700 IQ
Matrix adjustment, check and calibration
4 - 3
ba75651e03
08/2009
4.2
Matrix adjustment
4.2.1
General information on matrix adjustment
This procedure adjusts the value directly measured in the test sample
to an independently determined reference value ("lab value"). To
determine the reference values, a sample is taken from the measuring
solution and the relevant concentrations are measured (e.g.
photometrically).
First select for which of the installed combination electrodes the matrix
adjustment should be carried out. Based on this selection and the
electrode equipment, the sensor software determines the ion types for
which a reference measurement must be carried out. The menu guided
routine adapts itself correspondingly and informs you of all necessary
actions.
4.2.2
Carrying out the matrix adjustment
Main steps
Online help
A comfortable, menu guided routine guides you through the matrix
adjustment. All steps are easily and understandably explained on the
display. In addition, you can call up an
Online help
for each step. To do
so, use the arrow keys
<
STWX
>
or toggle switch
<
ST
>
to move the
highlighting to the
Online help
menu item and press
<OK>
. A help text
with detailed information on the relevant operating step appears on the
display. It provides, for example, important instructions on how to keep
the correct basic conditions. When
<OK>
is pressed once again, the
display returns to the current operating step.
Matrix adjustment with
electrode zeroing
After installing a new or different electrode you have to zero the
electrode to facilitate a long-term evaluation. Details on electrode
zeroing, see section 4.1.
Step 1:
Determination of all combination electrode voltages
("reference voltages"). The sensor is in the sample.
The procedure is started from the measured value display
with
<C>
. After completion the system returns to the
measured value display.
Step 2:
Sampling at the same place and time if possible and
determination of all relevant reference values
Step 3:
Entry of the measured reference values. This step is
started by pressing
<C>
again.
Summary of Contents for AmmoLyt Plus 700 IQ
Page 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Page 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Page 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Page 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Page 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...