21.3. Setting Sample Mass by Determining Mass of a Single Part
Enter
<3.2.PcS / 3.2.2.UUt>
submenu, set
<
S_S
>
value.
Enter
<PcS>
working mode (parts counting), blinking value of sample
quantity is displayed.
Press
key to select one of the following options:
10
Reference sample quantity: 10 pcs.
20
Reference sample quantity: 20 pcs.
50
Reference sample quantity: 50 pcs.
100
Reference sample quantity: 100 pcs.
0000
Custom reference sample quantity - enter the required value yourself.
Select respective option and press
key to confirm, first, text
<LoAd>
is displayed for 1 second, then the weighing window.
If the parts are to be weighed in a container, first put the container on
a weighing pan and tare it.
Load the weighing pan with declared amount of parts. When the indication
is stable (
pictogram is displayed), press
key to confirm
the mass.
Single part mass is calculated automatically, next quantity of parts (pcs)
is displayed.
Total weight value of all parts loaded onto the weighing pan
cannot be greater than the max capacity value.
Single part mass value must be equal or greater than 0.1
of the reading unit. Unless this condition is met,
the weighing device displays a message <Err Lo>.
In the course of parts quantity determination before
confirming the declared quantity value it is necessary to wait
for a stable measurement
.
22. WORKING MODE
– +/- CONTROL
+/- control is a working mode enabling you to enter checkweighing thresholds
values (
Min
,
Max
).