Steam Sterilizer Operation Manual
TANZO Classic Model Page of 46
28
Chapter 6 Operation Process
6.1 Switch On
Please connect the power before you start the operation. Turn on the power switch which
is located in the right panel of the sterilizer control box. When you turn on the power and
the sterilizer is not ready, the display interface shows the pot temperature, pressure,
showing flickering
“
LoAd
”
; If when you turn on the power and the sterilizer is ready,
flickering
”
LoAd
”
turns stable, the you can move to next step.
When t
he water shortage indicator “water” or astewater full indicator “drain”, one of them
or both of them are on, means the sterilizer is not ready. Only when the indicators are off
and the door is closed, the sterilizer is ready, the you can move to the nest step.
6.2 Distilled Water Adding
After you connecting to the power and select the corresponding program, if the “water”
indicator light illuminates, which means the water in the water storage tank is of its lowest
level and you have to add water in.
And
even you press“start/stop” button, it will not work either, so you need to add distilled
water in until the lowest required level can be reached.
You can fill in water on the top of the machine, as below (6-2-1). open the top cover and fill
in distilled wa
ter, stop filing water in when you hear “beep” warning sound.
(pic 6-2-1)