8.7 TRANSPORT OF DEVICE
When moving your K SERIES Cryostorage System to or from its current location to a transport vehicle
or new location for any reason, make sure to use a suitable forklift or pallet truck.
Always move your K SERIES Cryostorage System in an upright position
and in empty and defrosted condition, and without inserted inventory
control system.
Use straps to secure the refrigerator onto a palette.
Take particular care to lock all casters and use wheel chocks.
Fix flexible parts with low tack tapes or preferably bungee cords.
Protect the refrigerator with a blanket.
8.8 RETURN OF DEVICE
If it is necessary to return any part of the K SERIES Cryostorage System for repair,
maintenance or replacement, a Cryo
Product Return Information Form (see Appendix
C) must be filled in before returning the device to Worthington Industries.
Any K SERIES Cryostorage System received without this Cryo Product
Return Information will be returned to the sender.
It is your responsibility to assure that the goods are adequately packaged for return.
If necessary, consult Worthington Industries for shipping and packing instructions.
All refrigerators returned to Worthington Industries must be clean and disinfected before
return, as per section 8.5.
8.9 DISPOSAL OF DEVICE
The Worthington Industries K SERIES Cryostorage System has been designed and manufactured
with high quality materials and components, which can be recycled and reused.
Do not dispose of waste liquid nitrogen in-door
Dispose of waste liquid nitrogen out-of-doors where its cold temperature cannot
damage floors or driveways and where it will evaporate rapidly.
An outdoor pit filled with clean sand or gravel will evaporate liquid nitrogen safely and
quickly.