Care and Maintenance
45 / 62
Ethylene oxide sterilization is possible but not validated by the manufacturer.
9.5.2.9 Storage
After sterilization, the products in the sterilization packaging must be stored dry
and dust-free.
9.5.2.10 Reusability
The products can be reused up to 20 times, provided they are undamaged and un-
contaminated, and used with due care; any further use or the use of damaged and/
or contaminated products is the responsibility of the user.
If disregarded, any liability is excluded.
WARNING!
Signs of damage
Please check the reusable tube (NOT FOR SALE IN USA) for signs of damage after
sterilization and before use. Never use a tube showing signs of damage, especially
brittleness and perforation.
WARNING!
Contamination
Do not use the tube set if signs or indications of a contamination are detected.
CAUTION!
Maximal cleaning / reprocessing cycles
Do not clean / reprocess the reusable tube sets for more than 20 cycles as indicated
on the label.
CAUTION!
Check of the reprocessed tube set
After each cleaning / reprocessing, the cleaned / reprocessed tube set must be
checked for possible damages caused by the process before it is stored / used prop-
erly and professionally for the following use!