32
5.2
Calibration Function of Chemical Results
The reagent result can be adjusted according to an accurate quality control when some
analytes’ results of some reagent batches is deviated.
The specific steps are as follows:
1.
Open the analyser and choose
Settings
.
2. Select
Page Down
.
3.
Press
Modify Calibration
.
4.
Then change the “
from Date
” and
use the
Search
function.