Clinispin HORIZON  products are a  versatile  range  of centrifuges designed with 3 settings to process Chemistry, 
Coagulation or Platelet Poor Plasma and Urine specimens in the same unit. The maximum g-force of 2,000 xg makes 
Clinispin HORIZON compatible with most brands of tubes. Cycle settings can be changed to accommodate custom 
settings.

 

 
 

 

F

EATURES 

o

 

The first three (3) cycles are conveniently preset and labelled 

for your lab’s most common applications. Use the 

default cycles or customise them as needed.

 

 

o

 

If desired, the control panel can be temporarily locked on one cycle for error-free reproducibility.

 

o

 

Up to 10 cycles can be programmed for time, speed and braking and labelled with a custom name. Cycles can be 
programmed by g-force (RCF) rather than speed to facilitate matching validated cycles and manufacturer

s IFUs.

 

 

o

 

Lid lighting indicates the 

centrifuge’s

 status (ready, running, complete), informing the operator when tubes are 

ready for the analyser and preventing tubes from being left in the centrifuge longer than necessary (patent 
pending).

 

 

o

 

A traditional audible alert indicates the completion of the cycle. The audible alert can be muted.

 

o

 

Cool

Flow design prevents overheating of samples by using ambient air to keep specimens at room temperature.

 

o

 

The tube holders are fibre reinforced for high strength, durability and years of trouble-free use.

 

o

 

A clear lid permits safe observation of samples and optical calibration of speed.

 

o

 

The lid safety system prevents the centrifuge from operating unless the lid is closed and latched.

 

o

 

The lid safety system only allows entry into the centrifuge after the rotor has completely stopped.

 

o

 

The high-power brushless motor provides years of operation with no routine maintenance.

 

 

 

I

NTENDED 

U

SE 

General  purpose  laboratory  centrifuge  intended  for  safe  and  rapid  density  based  separation  of  fluids,  including 
physiologic fluids, in approved specimen receptacles for qualitative or quantitative test procedures. As a  general 
purpose  laboratory  centrifuge,  it  is  designed  to  also  run  other  approved  containers  filled  with  chemicals  (non- 
flammable, non-explosive, non-volatile and non-highly reactive only), environmental samples and other non- human 
body samples. This device is intended to be operated by properly trained personnel who have carefully read  the 
operating manual and are familiar with the function of the device. 

 

[Refer to the clinical laboratory method specified by the specimen receptacle manufacturer or established by the 
medical technology for the products applications.] 

 

 

W

ARRANTY 

Woodley Equipment Company warranties that this centrifuge is free from defects in workmanship and parts for 2 
years.

 

Summary of Contents for Clinispin Horizon 6 Flex

Page 1: ...Clinispin HORIZON 6 Flex 6 Flex FA Operator s Manual...

Page 2: ......

Page 3: ...1 Initial Setup 2 Quick Start 3 Single Cycle Lock 4 Preset Lock 4 Setting or Modifying a Cycle 4 Settings 5 Operation 5 Balancing Loads 6 Care and Preventative Maintenance 7 Cleaning and Disinfection...

Page 4: ...verheating of samples by using ambient air to keep specimens at room temperature o The tube holders are fibre reinforced for high strength durability and years of trouble freeuse o A clear lid permits...

Page 5: ...and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in which case the user will be...

Page 6: ...cycle displayed on the screen The lid must be closed Unlock Allows access into the rotor chamber by engaging the unlocking mechanism Entry is only possible when the rotor is stopped Stop Pressing the...

Page 7: ...es can again be edited or programmed SETTING OR MODIFYING A CYCLE Up to 10 cycles can be named programmed for time speed and braking and saved in memory Cycle Selection Press the up and down buttons n...

Page 8: ...ubes 1 850 xg N A 600 xg OPERATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise...

Page 9: ...e life of the centrifuge and produce better results Use the following rules when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired...

Page 10: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Page 11: ...er with a tool to remove them o Inspect the rotor tube holders or buckets for damage o If the tube holders or buckets have any damage evenslight safely dispose of them and replace them o If the rotor...

Page 12: ...CK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label and use apen to manually disengage the locking mechanism...

Page 13: ...2 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 220 Watts 220 Watts Centrifuge Motor H P Brushless H P Brushless Max G force 2 000 xg 1 600 xg 1 850 xg Max Speed...

Page 14: ...e of their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipm...

Reviews: