
Measurement Modes in the Main Menu
48
Fig. 54: Presettings (Volume test)
Select "Start measurement" and press "Start".
In the display the request to add/remove the sam-
ple volume will appear.
Take the sample volume with the syringe or
the soot test pump.
Fig. 55: Volume test
When the value V keeps stable, press "OK to
save the data.
In the main menu a tick mark beside "Volume" will
indicate that the volume test already has been
done.
V
TST
= Volume
Test
P
T
= Pressure
Tube
P
amb
= Ambient Pressure
6.9
Graph / Record
Fig. 56: "Record" screen
The "graph / record" mode is used to start a per-
manent data recording. A graph of the records will
appear on screen.
The user can modify the log interval, the number
of measurements, the stop date and the stop time
as follows:
Go to the digit by pressing the left-/right-key
and increase or reduce the value by pressing
the up- and down-keys.
The display can be switched off during the data
recording.
NOTE!
The display only will switch off between the
measuring cycles, if a record interval of at least
20 seconds has been set.
Select "Start recording" and press "Start"