Use
38 / 96
17.10.2019
5.2.5
Programming a quick test
Test pressure
ü
The ProofMaster shows the home screen.
a
a) Press
.
b
e
g
ð
The Quick Select screen appears.
b) Tap
Pressure
.
ð
The selection of the test pressure values appears.
c) Press the desired test pressure or press
Manual
.
ð
The test pressure is programmed or a number field appears
when
Manual
is selected.
d) For an individual test pressure, enter the desired test pressure
value between -0.7 bar and 10.0 bar and then press
to
confirm the entry.
ð
The individual test pressure is programmed.
e) Press
Watch size
to program the size of the watch.
ð
The selection of the watch sizes appears.
f) Press the value corresponding to the size of the watch.
g) If necessary, press
Vacuum test
to perform a vacuum test before
the pressure test.
Info
The pressure for the vacuum test is fixed.
The pressure for the vacuum test is -0.3 bar and cannot be changed.
ð
The
Vacuum test
button is displayed in blue.