![Winmate M270TF-SDI User Manual Download Page 16](http://html1.mh-extra.com/html/winmate/m270tf-sdi/m270tf-sdi_user-manual_987007016.webp)
15
4K UHD 12G-SDI LCD Monitor User Manual
3.1.5 Moisture condensation
If the unit is suddenly taken from a cold to a warm location, or if ambient temperature suddenly rises,
moisture may form on the outer surface of the unit and/or inside of the unit. This is known as condensation.
If condensation occurs, turn off the unit and wait until the condensation clears before operating the unit.
Operating the unit while condensation is present may damage the unit.
3.1.6 Cleaning before cleaning
Be sure to disconnect the AC power cord from the AC outlet.
3.1.7 Cleaning the monitor
A material that withstands disinfection is used for the front protection plate of the medical use LCD monitor.
The protection plate surface is specially treated to reduce reflection of light. When solvents such as benzene
or thinner, or acid, alkaline or abrasive detergent or chemical cleaning cloth are used for the protection plate
surface/monitor surface, the performance of the monitor may be impaired or the finish of the surface may be
damaged. Take care with respect to the following:
Clean the protection plate surface/monitor surface with a 50 to 70 v/v% concentration of isopropyl
alcohol or a 76.9 to 81.4 v/v% concentration of ethanol using a swab method. Wipe the protection
plate surface gently (wipe using less than 1 N force).
Stubborn stains may be removed with a soft cloth such as a cleaning cloth lightly dampened with
mild detergent solution using a swab method and then clean using the above chemical solution.
Never use solvents such as benzene or thinner, or acid, alkaline or abrasive detergent, or chemical
cleaning cloth for cleaning or disinfection, as they will damage the protection monitor surface.
Do not use unnecessary force to rub the protection plate surface/monitor surface with a stained cloth.
The protection plate surface/monitor surface may be scratched.
Do not keep the protection plate surface/monitor surface in contact with a rubber or vinyl resin
product for a long period of time. The finish of the surface may deteriorate.
3.1.8 Flat surface for better maintenance
The design allows the user to easily wipe liquids and gel off the LCD panel and control buttons
– ensuring
a high standard of disinfection and cleanliness.
3.1.9 Repacking
Do not throw away the carton and packing materials. They can be used again to repack monitor.
If you have any questions about this unit, contact your authorized dealer.
3.1.10 Disposal of the unit
Do not dispose of the unit with general waste. Do not include the monitor with household waste.