Power Level
Accuracy
Test Proce-
dure
Power level accuracy is checked by stepping the
power down in 1 dB increments from its maximum
rated power level.
1. Set up the 681XXA as follows:
a. Reset the instrument by pressing
SYSTEM
,
then
Reset
. The CW Menu is displayed.
b. Press
Edit F1
to open the current frequency
parameter for editing.
c. Set F1 to the CW frequency noted on the test
record for the model being tested.
d. Press
Edit L1
to open the current power level
parameter for editing.
e. Set L1 to the power level noted on the test re-
cord.
2. Measure the output power level with the Power
Meter and record the reading on the test record.
3. Verify that the Power Meter reading meets the
specifications stated on the test record.
4. Set L1 to the next test power level. Record the
Power Meter reading on the test record.
5. Repeat step 4 for the other levels listed on the
test record for the current CW frequency.
6. Repeat steps 1 thru 5 for all CW frequencies
listed on the test record.
OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
5-10
681XXA OM