background image

28

Guidance and Manufacturer’s Declaration -  

Electromagnetic Immunity

LimbLogic is intended for use in the electromagnetic environment 
specified below. The customer or the user of the LimbLogic should 
assure it is used in such an environment.

Immunity  

test

IEC 60601 

test level

Compliance 

level

Electromagnetic  

environment - 

guidance

Electrostatic 
discharge 
(ESD)       
IEC 61000-4-2

± 6 kV 
contact

± 8 kV air

± 6 kV  
contact

± 8 kV Air

Floors should be 
wood, concrete, or 
ceramic tile. If floors 
are covered with 
synthetic material, 
the relative humidity 
should be at least 
30%.

Electrical fast 
transient/
burst  
IEC 61000-4-4

± 2 kV for 
power  
supply lines

± 1 kV on  
input/ 
output lines

± 2 kV on 
power  
supply lines

± 1 kV on  
Input/ 
output lines

Mains power quality 
should be that of a 
typical commercial 
or hospital  
environment.

Surge  
IEC 61000-4-5

± 1 kV  
differential 
mode

± 2 kV  
common 
mode

± 1 kV  
differential 
mode

± 2 kV  
common 
mode

Mains power quality 
should be that of a 
typical commercial 
or hospital  
environment.

Voltage dips, 
short interrup-
tions and volt-
age variations 
on power 
supply input  
lines  
IEC 61000-4-11

< 5% U

T

 

(>95 % dip  
in 

U

T

) for  

0,5 cycle

40% 

U

T

 

(60% dip in 

U

T

) for  

5 cycles

70% 

U

T

 

(30% dip  
in 

U

T

) for  

25 cycles

< 5% 

U

T

 

(>95 % dip  
in 

U

T

) for  

5 sec

< 5 % U

T

(>95 % dip  
in 

U

T

) for  

0,5 cycle

40% 

U

T

 

(60% dip  
in 

U

T

) for  

5 cycles

70% 

U

T

 

(30% dip  
in 

U

T

 for  

25 cycles

< 5% 

U

T

 

(>95 % dip  
in 

U

T

) for  

5 sec

Mains power quality 
should be that of a 
typical commercial 
or hospital  
environment. If 
the user of the 
LimbLogic requires 
continued  
operation during 
mains interruptions, 
it is recommended 
that the LimbLogic 
be powered from 
an uninterruptible 
power supply or a 
battery.

Power  
frequency 

(50/60 Hz)
magnetic field 
IEC 61000-4-8

3 A/m

3 A/m

Power frequency 
magnetic fields 
should be at levels 
characteristic of  
a typical location 
in a typical com-
mercial or hospital 
environment.

NOTE  U

T

 is the a.c. mains voltage prior to application of the test level.

Technical Information

Summary of Contents for LimbLogic Series

Page 1: ...Quick Start Guide Quantity 2 for bilateral kit This document provides instruction for the prosthetist in the installation and use of the system A separate document LimbLogic Patient Instructions is i...

Page 2: ...ventilated area at a temperature between 5 C 41 F and 33 C 90 F Failure to do so may result in prolonged or incomplete charging of the Vacuum Pump and may damage the Vacuum Pump A Vacuum Pump enclose...

Page 3: ...the following two conditions 1 this device may not cause interference and 2 the user of the device must accept any interference including interference that may cause undesired operation Remarque Le pr...

Page 4: ...urning On the Vacuum Pump 10 Turning Off the Vacuum Pump 11 Resetting the Vacuum Pump 11 Turning On the Fob 11 Changing Vacuum Set Point 12 Changing Pump State 12 Bilateral Indicators 13 Changing the...

Page 5: ...n be pried open using a small flat blade screw driver then snapped back together starting at the battery end and working back toward the opposite end The fob battery could cause serious injury or deat...

Page 6: ...r button until the yellow digits flash in a chasing pattern and then display the current vacuum level The pump is now Active and will operate until the preselected vacuum level has been reached 3 To e...

Page 7: ...the pump stack the second gasket on top of the first gasket before attaching the pump to the socket You should be able to turn the pump upside down as shown without the gasket falling off If the gask...

Page 8: ...set at 20 but if those settings are not suitable for your patient you may adjust them as desired The User Maximum cannot be set higher than the system maximum of 20 The optimum vacuum setting is diffe...

Page 9: ...page 11 or press the on off button on the Vacuum Pump The pump will operate until the preselected vacuum level is reached Note Your pump can be used while charging If the pump has just been unplugged...

Page 10: ...ton on the Vacuum Pump one time The Vacuum Pump will beep once Before turning on the fob as described below wait 10 seconds for the pump to activate Overview on off button on off button Your battery w...

Page 11: ...empt to connect to the second pump using the same steps as above If communication cannot be established with either pump the fob will shut down RESETTING THE VACUUM PUMP It is possible for a large sta...

Page 12: ...le the setting is being applied If the Enter button is not pressed the settings will not be applied and the fob will shut off due to inactivity in 60 seconds CHANGING PUMP STATE Make sure the fob is o...

Page 13: ...ne of the two white indicators at the top of the fob will illuminate to show which pump left or right is being controlled To switch the fob from controlling one pump to the other press and hold the Up...

Page 14: ...ximum The User Maximum cannot be set higher than 20 Press the Enter button to apply the new User Maximum setting and advance to the next setting The display will alternate between two quick flashes of...

Page 15: ...is charging The level of charge is indicated as follows Solid green Vacuum Pump battery is between 50 and 100 charged Flashing green Vacuum Pump battery is between 30 and 50 charged Solid red Vacuum P...

Page 16: ...nium Pyramid FND 227014 Capscrews Hose Barb connects to external hose on pump VACUUM PLATE PYRAMID The Vacuum Plate LLV 01043 and Vacuum Pyramid LLV 01044 can be used for attaching the Distal Mount pu...

Page 17: ...t the existing tube just past the Muffler 3 Connect the Exhaust Tube Fitting to the section of tube that does not have the Muffler The pump will function normally but at a slightly higher noise level...

Page 18: ...ally stalled Refer to the Pump Stall error code information on page 21 Pump battery does not maintain charge for a full day Ensure that the system is properly sealed Ensure that the battery has been c...

Page 19: ...r at least 30 minutes and try again The pump may have been exposed to a large static charge Press and hold the on off button on the pump until you hear a beep approximately 12 seconds Does the patient...

Page 20: ...e Fitting from the pump housing Do not lose the o ring from the fitting Install the drain tube and flush the system with alcohol The pump beeps but does not run The Muffler on the Exhaust Tube may be...

Page 21: ...rature to return to operating range then reset the pump by pressing and holding the pump s on off button for approximately 12 seconds until you hear a beep 03 Motion fail Motion sensor failure Return...

Page 22: ...um Pump cycles more than once every 20 seconds the Vacuum Pump has a leak If the vacuum level holds steady the pump is functioning correctly 7 Remove the Diagnostic Disc or Sheet and attach the pump a...

Page 23: ...d abusive activities are defined as activities such as skydiving karate and judo activities that could result in injury to an individual s natural limbs and activities that expose the prosthesis to co...

Page 24: ...This product complies with applicable standards for sale and use in EC countries This device operates on up to 7 volts of AC and DC current Type BF applied part On Off button Push Push style This prod...

Page 25: ...e LimbLogic vacuum system has been designed for and clinically tested as a suspension aid It has not been clinically tested for wound healing or other uses WillowWood does not currently support the us...

Page 26: ...IC 7475A 21011 This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept an...

Page 27: ...technicien exp ri ment dans le domaine radio TV ELECTROMAGNETIC COMPATIBILITY Medical electrical equipment needs special precautions regarding electromagnetic compatibility EMC and needs to be install...

Page 28: ...mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrup tions and volt age variations on power supp...

Page 29: ...n elec tromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MH...

Page 30: ...distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitte...

Page 31: ...ortation or short term storage between uses Temperature 25 C to 70 C Humidity 15 to 93 non condensing Altitude 1250 to 20000 feet 380 to 6000 meters Pressure 1060 to 470 hPA Long term storage Temperat...

Page 32: ...Mt Sterling OH 43143 phone 740 869 3377 800 848 4930 fax 740 869 4374 www willowwoodco com PN 2025 Y 28 AUG 2015 Ohio Willow Wood Company B V Keizersgracht 62 64 1015 CS Amsterdam The Netherlands Pate...

Reviews: