background image

22

Overview

DIAGNOSTIC DISC OR SHEET

The Diagnostic Disc and Sheet are included with LimbLogic 
Systems. These are used to evaluate the internal seal of the 
pump. The Disc is used with the distal mounted design, and 
the Sheet is used with the side mounted design. Do not use the 
sealing Disc with a Side Mounted pump. The Disc and Sheet are 
used to evaluate the pump’s seal integrity as follows:

1. 

If you are using a new disc with a Distal Mount pump,  
remove the protective tape from the Diagnostic Disc.

2.  Place the Sealing Gasket in the recess around the inlet to 

the pump.

3.   Place the Disc (if using a Distal Mount pump) or the Sheet 

(if using a Side Mount pump) over the inlet and the gasket.

4.  Gently bolt the pump back onto the socket to ensure a 

tight seal. Note: If evaluating a Side Mount pump and a 
sealing plate, make sure that the plate is aligned between 
the mounting bolts before gently bolting on the pump. 

5.  Note the Vacuum Pump’s vacuum level and range settings 

and then set the vacuum level to 20 and the range to 4.

6.   Start the pump and watch the display on the fob:

•  If the Vacuum Pump cycles more than once every 20 

seconds, the Vacuum Pump has a leak.*

•  If the vacuum level holds steady, the pump is functioning 

correctly.

7. 

Remove the Diagnostic Disc or Sheet and attach the pump 
and gasket for normal operation of the system.

8.  Return the Vacuum Pump to its original vacuum and range 

settings.

*All pneumatic systems leak. The volume of air trapped behind the  
Diagnostic Disc or Sheet is typically 10 to 20 times smaller that the  
volume of a typical socket. This results in a very sensitive measure of 
leakage. A Vacuum Pump that passes this test will only use a small  
percentage of its battery life to accommodate Vacuum Pump leakage.

Diagnostic Disc

Sealing Gasket

LimbLogic 

Vacuum Pump

Capscrews

OPERATION AT HIGH ALTITUDE OR LOW PRESSURE

LimbLogic uses air pressure to hold on a prosthesis. There is 
less air pressure at higher altitudes. LimbLogic will function 
normally at up to 1700 meters (5500 feet) of altitude. However, 
the pressure available to hold on your prosthesis may be lower 
if you are over 1700 meters (5500 feet). If this happens, your 
pump will indicate that it has a leak and you should lower your 
vacuum setting. A setting of 17 can usually be used for altitudes 
of up to 3000 meters (10000 feet). 

Diagnostic 

Sheet 

Sealing  

Gasket

Side Mount 

Vacuum 

Pump

Summary of Contents for LimbLogic Series

Page 1: ...Quick Start Guide Quantity 2 for bilateral kit This document provides instruction for the prosthetist in the installation and use of the system A separate document LimbLogic Patient Instructions is i...

Page 2: ...ventilated area at a temperature between 5 C 41 F and 33 C 90 F Failure to do so may result in prolonged or incomplete charging of the Vacuum Pump and may damage the Vacuum Pump A Vacuum Pump enclose...

Page 3: ...the following two conditions 1 this device may not cause interference and 2 the user of the device must accept any interference including interference that may cause undesired operation Remarque Le pr...

Page 4: ...urning On the Vacuum Pump 10 Turning Off the Vacuum Pump 11 Resetting the Vacuum Pump 11 Turning On the Fob 11 Changing Vacuum Set Point 12 Changing Pump State 12 Bilateral Indicators 13 Changing the...

Page 5: ...n be pried open using a small flat blade screw driver then snapped back together starting at the battery end and working back toward the opposite end The fob battery could cause serious injury or deat...

Page 6: ...r button until the yellow digits flash in a chasing pattern and then display the current vacuum level The pump is now Active and will operate until the preselected vacuum level has been reached 3 To e...

Page 7: ...the pump stack the second gasket on top of the first gasket before attaching the pump to the socket You should be able to turn the pump upside down as shown without the gasket falling off If the gask...

Page 8: ...set at 20 but if those settings are not suitable for your patient you may adjust them as desired The User Maximum cannot be set higher than the system maximum of 20 The optimum vacuum setting is diffe...

Page 9: ...page 11 or press the on off button on the Vacuum Pump The pump will operate until the preselected vacuum level is reached Note Your pump can be used while charging If the pump has just been unplugged...

Page 10: ...ton on the Vacuum Pump one time The Vacuum Pump will beep once Before turning on the fob as described below wait 10 seconds for the pump to activate Overview on off button on off button Your battery w...

Page 11: ...empt to connect to the second pump using the same steps as above If communication cannot be established with either pump the fob will shut down RESETTING THE VACUUM PUMP It is possible for a large sta...

Page 12: ...le the setting is being applied If the Enter button is not pressed the settings will not be applied and the fob will shut off due to inactivity in 60 seconds CHANGING PUMP STATE Make sure the fob is o...

Page 13: ...ne of the two white indicators at the top of the fob will illuminate to show which pump left or right is being controlled To switch the fob from controlling one pump to the other press and hold the Up...

Page 14: ...ximum The User Maximum cannot be set higher than 20 Press the Enter button to apply the new User Maximum setting and advance to the next setting The display will alternate between two quick flashes of...

Page 15: ...is charging The level of charge is indicated as follows Solid green Vacuum Pump battery is between 50 and 100 charged Flashing green Vacuum Pump battery is between 30 and 50 charged Solid red Vacuum P...

Page 16: ...nium Pyramid FND 227014 Capscrews Hose Barb connects to external hose on pump VACUUM PLATE PYRAMID The Vacuum Plate LLV 01043 and Vacuum Pyramid LLV 01044 can be used for attaching the Distal Mount pu...

Page 17: ...t the existing tube just past the Muffler 3 Connect the Exhaust Tube Fitting to the section of tube that does not have the Muffler The pump will function normally but at a slightly higher noise level...

Page 18: ...ally stalled Refer to the Pump Stall error code information on page 21 Pump battery does not maintain charge for a full day Ensure that the system is properly sealed Ensure that the battery has been c...

Page 19: ...r at least 30 minutes and try again The pump may have been exposed to a large static charge Press and hold the on off button on the pump until you hear a beep approximately 12 seconds Does the patient...

Page 20: ...e Fitting from the pump housing Do not lose the o ring from the fitting Install the drain tube and flush the system with alcohol The pump beeps but does not run The Muffler on the Exhaust Tube may be...

Page 21: ...rature to return to operating range then reset the pump by pressing and holding the pump s on off button for approximately 12 seconds until you hear a beep 03 Motion fail Motion sensor failure Return...

Page 22: ...um Pump cycles more than once every 20 seconds the Vacuum Pump has a leak If the vacuum level holds steady the pump is functioning correctly 7 Remove the Diagnostic Disc or Sheet and attach the pump a...

Page 23: ...d abusive activities are defined as activities such as skydiving karate and judo activities that could result in injury to an individual s natural limbs and activities that expose the prosthesis to co...

Page 24: ...This product complies with applicable standards for sale and use in EC countries This device operates on up to 7 volts of AC and DC current Type BF applied part On Off button Push Push style This prod...

Page 25: ...e LimbLogic vacuum system has been designed for and clinically tested as a suspension aid It has not been clinically tested for wound healing or other uses WillowWood does not currently support the us...

Page 26: ...IC 7475A 21011 This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept an...

Page 27: ...technicien exp ri ment dans le domaine radio TV ELECTROMAGNETIC COMPATIBILITY Medical electrical equipment needs special precautions regarding electromagnetic compatibility EMC and needs to be install...

Page 28: ...mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrup tions and volt age variations on power supp...

Page 29: ...n elec tromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MH...

Page 30: ...distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitte...

Page 31: ...ortation or short term storage between uses Temperature 25 C to 70 C Humidity 15 to 93 non condensing Altitude 1250 to 20000 feet 380 to 6000 meters Pressure 1060 to 470 hPA Long term storage Temperat...

Page 32: ...Mt Sterling OH 43143 phone 740 869 3377 800 848 4930 fax 740 869 4374 www willowwoodco com PN 2025 Y 28 AUG 2015 Ohio Willow Wood Company B V Keizersgracht 62 64 1015 CS Amsterdam The Netherlands Pate...

Reviews: