Scroll along the black
selection bar or increase/decrease
the selected parameter�
KD:
Quit edit mode�
KD:
Select edit (yellow), change with
plus and minus buttons and confirm�
7.2.1 Reducing condensation in
the inspiration tube
To reduce condensation in the
inspiration tube, it is recommended that
the temperature difference between
the patient-side setpoint temperature
and the setpoint temperature of the
humidification chamber be increased�
7.2.2 Adjusting humidity
The humidity near the patient increases
when the setpoint temperature of the
chamber is increased�
The humidity near the patient decreases
when the setpoint temperature of the
humidification chamber is reduced�
7.2.3 Reducing condensation in
the expiration tube
To reduce condensation in the
expiration tube, it is recommended
that the heat output of the expiratory
heating wire be increased�
Operation in edit mode(free)