WIKA operating instructions, model CTR3000
29
14191761.01 05/2019 EN/DE
6.4.1.8 Peak “clear” function
The peak function (Min./Max.) shows the maximum and minimum value for the actual recorded measurement in the unit
displayed in the button
10
.
The
Peak
field
17
shows in the first line peak
Max.
and in the second line peak
Min.
By pressing the button
12
[Clear]
the values will be deleted and the values are refreshed.
Resolution fixed at 4 respectively 5 decimal places, only depending on unit selection.
■
Temperature selected → 4 places
■
Base value selected → 5 places
So only a change of the unit causes change of peak values.
6.4.1.9 Information bar
The information bar, field
16
, displays relevant measurements which help you to have an easier calibration documentation.
Values
Number of measurements
Setting in
[Settings] / [Display] / [Average value]
Average
Average value over the last xy values; displayed in the selected unit on the main screen
Stability
Shows the standard deviation
Difference
Displays the difference between the reference (first probe marked as reference) and the test item (all
other channels); displayed in the unit of the test item (only shown in scan mode when the selected view is
calibration, please refer to chapter 6.4.4 “Application [Scan]”)
Information bar has the same colour as the corresponding channel selection on the main screen (blue, green, red
and orange). If channels are displayed in a list or no channel relevant information is displayed the information bar
will be displayed in grey.
12
10
17
16
6. Commissioning, operation