8. Remote Operation
58
WIKA Operating Instructions High Pressure Controller, Model CPC8000-H
GB
1
4
2
3
5
9
9
3
.0
0
/0
6
/2
0
1
7
/GB
8.2 Mensor Command Set
The following table lists all of the current CPC8000 commands and queries.
Optional emulation modes are available in which a CPC8000 can emulate remote functions of different
brands of pressure gauges. Please contact WIKA/Mensor for more details.
Table 6 Mensor command set
Item Command or
Query
Data
Response/Function
1.
?
none
Returns data per the
current output form.
2.
ACQUIRE?
15 char string.
Example:
Acquire? Test_stand_1
Returns:
<sp>(yes or no),CCC...
CCC<cr><lf>
This command is used
when multiple computers
would like to control the
instrument. Yes if
acquisition is successful.
No if instrument is being
controlled with another
computer. CCC... = name
of controlling computer.
See: RELEASE? and
UNLOCK
3.
ADDRESS
0-31
Sets the GPIB Address
4.
ADDRESS?
<sp> xx <cr><lf>
Returns the GPIB
Address
5.
ASSET_TAG
16 char string
General purpose string for
customer use.
6.
ASSET_TAG?
<sp>ssssssssssssssss<cr><lf>
Return customer asset tag
string.
7.
AUTORANGE
ON or OFF
Sets whether the
autorange function is
enabled or disabled.
8.
AUTORANGE?
<sp>(ON or OFF)<cr><lf>
Returns whether the
autorange function is
enabled or disabled.
9.
AUTOZERO
none
Re-zero all the ranges.
These adjustments are
not password protected
and are not saved through
power cycles. This
command takes
approximately 60 seconds
but may take longer
depending on the time to
become stable.
10.
AUTOZERO?
S,T,X,X
Returns autozero data
where S represents state
(0 = complete, 1 = local
autozero, 2 = remote
autozero), T represents
the estimated time to
complete in seconds, and
x is a (0) character since
this data location is not
used at this time.