7. Operation
WIKA Operating Instruction High Pressure Controller, Model CPC8000-H
41
GB
1
42
3
59
9
3
.0
0/
0
6/2
0
17
/GB
7.3 Setup
After the instrument has been powered on and the main screen is displayed, press the Setup [
]
button on the sidebar to setup parameters or make changes to the instrument settings.
Figure 7-22 Location of the Setup button
The following picture is a menu tree that shows the general layout of the setup menus. The bullet
points in blue represent screens that require the first level password to view; those in red require a
second level password to view.
Figure 7-23 Setup Screen Tabs
7.3.1 General Tab
Language
The languages that are available in the CPC8000 are shown by pressing Setup [
], then press the
[General] tab, and then the [Language] button. Note that the active language is indicated on the right
side of the [Language] button. When the [Language] button is pressed, the sidebar will show the
languages that are available. Pressing the desired language on the sidebar will make that language
active on every screen in the CPC8000.
Figure 7-24 Language
Secondary and Tertiary Display
The Secondary and Tertiary display areas provide a place to indicate extra data that may be desired
during testing or calibration. In the [General] tab, there is a selection button labeled [Secondary
Display] and one labeled [Tertiary Display]. Clicking either of these will change to a sidebar to indicate
the choices that are available to each of these display areas. Pressing any one of these choices will
make that choice active in its respective display area on the main menu.
Setup [
] button