2.6.3 Output Lead Electrode
This section allows the user to select which lead electrode the output is connected to (e.g.
terminal P1 in the IEC 60601-2-25, Figure 201.106). Unselected electrodes are connected to
the system ground (terminal P2 in Figure 201.106).
More than one lead electrode may be selected. For example, if it is desired to have Lead I
and Lead II have a positive indication, LA and LL can be selected.
2.6.4 Output Summary
The output waveform and related settings shall display in this section. User can have a
better idea of the waveform details.
User can also click “Save as Script “to
save the waveform settings as an individual file. The
script file can be loaded in the auto sequence function. Please refer to section 2.7.6 for
details.
2.6.5 Output Waveform Display
The output display provides an image similar to that provided by ECGs. The sensitivity of the
display range may be set at 5mm/mV, 10mm/mV or 20mm/mV to cover the full range of
waveforms offered by the system. The time rate is fixed.
The output display uses the same data as used in the DAC output and serves as a cross check
of the selected waveform, and also allows the user to view the original waveform as filters in
the ECG device under test can substantially alter the waveform. The pacing signal is shown in
red color.