![WHALETEQ DFS360 User Manual Download Page 23](http://html.mh-extra.com/html/whaleteq/dfs360/dfs360_user-manual_967654023.webp)
DFS360
|
User Manual
22
Figure 16: Check Desired Test Results for Analysis
(2) After checking desired test results, click Reproducibility button to
bring up the Reproducibility Analysis window for the analysis result. The
upper section of the window shows shock waveforms; the lower section
shows energy (in Joules). Users can drag the orange frame on the lower
section to select a designated range of test results to superimpose their
shock waveforms on the upper section.
At the upper-right corner on the lower section (the green circle on
the figure below), the auto-calculated mean value and standard deviation
show as a reference for checking whether the joules of each test result is
standard.
By superimposing waveforms, users can confirm the consistency of
shock waveforms of multiple DUTs; or whether the multiple test results of a
single DUT conform to the original design.