AECG100 | User Manual
Specification
9.1
ECG Test Mode
Table 4: ECG Test Mode Specification
Parameter
Specification
Main Output Voltage
Resolution 2.5
μ
V (DAC resolution)
Accuracy
±
1% for amplitudes of
0.5mVpp or higher
Frequency/ Pulse repetition
rate
Accuracy
±1 %
Pulse duration/ Timing
Accuracy
±
1ms
Pacing pulse amplitude
Range
-1000mV ~ +1000mV
Accuracy
±
2mV pulse:
±
1%
>2mV or <-2mV pulse:
±
10%
Pacing pulse width
Range
0.1ms ~ 2ms
Resolution 0.1ms
Accuracy
±
5
μ
s
Pacing pulse characteristics
Rise/ fall
time
5μs
Overshoot <1%
Settling
time
<1%
Sample rate
Rate
50 ~ 40K (Normal/
Standalone);
10K (Raw data mode)
Accuracy
Depend on system clock
accuracy
DC offset
(fixed, noise free, sourced
from internal super
capacitor)
Range
-300mV, +300mV
Accuracy
±
1%
DC offset
Range
-500mV ~ +500mV