![WHALETEQ AECG100 User Manual Download Page 73](http://html.mh-extra.com/html/whaleteq/aecg100/aecg100_user-manual_967648073.webp)
73
AECG100 | User Manual
Figure 88: DC Voltage Verification (>300mV)
Troubleshooting
(1)
If the PPG module is accidentally separated from AECG100 main
console unit in operation, please follow the below steps to have
the test system work properly.
Step1: Turn off the power switch on AECG100 main console unit
and assure the power indicator is off.
Step2: Re-connect the PPG module to the AECG100 main console.
Step3: Turn on the power switch on AECG100 main console unit
and assure the power indicator is on.
(2)
When Red LED of the power indicator remains on:
If Red LED remains on, the AECG100 system is crashed. Please try
turning off and then turning on the AECG100 test system.
(3)
No LED display at the power indicator:
For no LED display, either the system is crashed or no power is
supplied to the test system. Please try to unplug and re-plug in the