66
AECG100 | User Manual
from the beginning.
(9)
Loop:
To replay the selected test sequence repeatedly.
(10)
Total:
Show the total playback time of the set sequence.
4.8
Supported Raw Data File Format
AECG100 support Text file (*.txt) only and file content describes as
below:
File Format
:
[sampling frequency]
[number of samples per signal]
[number of signals]
[signal description (signal-1)],[signal description (signal-2)], ...
[sample data-1 (signal-1)],[sample data-1 (signal-2)], ...
[sample data-2 (signal-1)],[sample data-2 (signal-2)], ...
...
[sample data-N (signal-1)],[sample data-N (signal-2)], ...
Description:
sample frequency
Input the frequency of sample data
number of samples per signal
The limitation is related to the available memory size of PC system. If
PC system has available memory for 1Gb, the maximum number of
single channel sample data would be 131,072.
number of signals
This number must be greater than 1.
signal description
The description text for the signal.
sample data
The raw data unit is mV.