48
AECG100 | User Manual
Note:
If the transmission SpO
2
module has ECG function, the PWTT mode
operation is the same as the above.
4.5
Auto Test SpO
2
Mode
This mode is only applicable to the PPG-2TF-660. The PPG-2TF-660 can be
used only after it is connected to the AECG100.
When the PPG-2TF-660 detects the DUT, the system will
identify the
parameter settings such as Trigger Level and DC value of R/IR, and record
Trigger Level to SpO
2
and PPG tabs.
If the Auto Trigger Level reminder shows, please check the IR status, or
whether the DUT connects to the PPG-2TF-660 and is in a proper measuring
position. After confirming the aforementioned
situations, click ‘’
OK
’’ to
continue.
Figure 54: Auto Trigger Level reminder
Auto Test SpO
2
Mode tab can be separated into 2 sections. The upper section
is for waveform selection and test parameter setting; the lower section is for
waveform playback display and setting. This mode can detect the dynamic
light intensity function of the DUT, so the PPG-2TF-660 can send
corresponding signals as the same intensity to the DUT according to the
change in light intensity.